Return to Sterilizer Selection Screen

Benchtop Sterilizers

Classic Models

UMB420BT (33l)

UMB430BT (43l)

UMB440BT (63l)


Autofill Models

UMB220BT (33l)

UMB230BT (43l)

UMB240BT (63l)


PDF Brochure

Top Loading

Classic Models

UMA440BT (63l)

UMB240BT (63l)


UMB230BT - 43L 'Autofill' Sterilizer

Suitable for Wastewater, Food Safety, Veterinary and Microbiology applications.

Astell’s front loading Benchtop autoclaves require no plumbing and operates from a standard single phase power supply. Many customers choose this model not only for its small footprint, but also because it requires no services other than a standard power supply.

The ‘Autofill’ version has an integral reservoir. Once filled, this provides water for up to 15 cycles. The reservoir also acts as a collection vessel for condensate and the water that is automatically returned from the chamber at the end of each cycle.

This model is suitable for sterilizing:

  • Liquids (media)

  • Lab waste

  • Glassware

  • Lab instruments

This model features

  • A fully programmable touch screen color controller

  • An electro polished stainless steel chamber

  • A base shelf within the chamber

  • An ultra quick Swiftlock door mechanism

  • A self contained design with no installation required - just fill it with water and plug it in and the machine is ready to be used.

  • Angled chamber orientation, ensuring a greatly reduced water requirement every sterilizaion cycle.

Markets Astell Sells Autoclaves in

Technical details - UMB230BT

The 'Autofill' version of Benchtop autoclaves uses Astell’s unique recirculation system - meaning that there is no need for the autoclave to be connected to any services other than a standard electric power supply. The autofill version of the Benchtop range differs from the 'Classic' version as it has an integral reservoir built into the inside of the unit. The user simply has to top up this internal tank via a removable screw lid, and a single fill will last for approximately 15 - 20 cycles.

At the start of each cycle the chamber is automatically filled from the reservoir, and following sterilization the autoclave then automatically drains any excess water from the chamber and returns it to the internal tank. This saves the user having to top up the autoclave after every cycle, and also ensures that no provision for condensate collection is necessary.

Criteria

Details

Model Number/Type:

UMB230BT - Autofill Benchtop Autoclave

Steam Source:

Heaters in the Chamber

Capacity:

43 litres

Power Supply:

208v/220v, 60Hz

Chamber dia x depth

350mm x 460mm

Overall Dims (w x h x d)

520mm x 540mm x 780mm

500ml Duran Capacity

14

1000ml Duran Capacity

11

2000ml Duran Capacity

3

Integral Data Printer - REF: AAR130

This printer provides a permanent and traceable record of: time, temperature, pressure, the batch number and the cycle name. There is also provision for operator signature; the printer also provides reports of cycle settings and servicing information.

Astell Scientific Printer

Load Sensed Process timing - REF: AAR014

Load Sensed Process Timing allows the sterilization cycle to be controlled via the temperature achieved in the centre of the load. The operation of load sensed process timing is controlled via the ‘Logi’ touch screen controller. A ‘wandering’ probe, situated within the chamber is inserted into the load, or load simulator, and initiates the sterilization period once the probe reaches the programmed threshold temperature.

Load Sensed Process Timing

RS232 Interface - REF: AAR120

The addition of the RS232 option allows cycle progress to be monitored (via HyperTerminal) on an external computer equipped with suitable software.

RS232

Ethernet Interface - REF: AAR122

The addition of the Ethernet interface allows the visual display of the colour touch screen controller on another device. This works by using a VNC Viewer (not supplied) and a PC or other hardware capable of running VNC.

RJ45

IQ/OQ Documentation - REF: IQ/OQ

The IQ/OQ documents package provides the user with blank documents for them to complete. The IQ section includes details of calibration equipment, an order Acknowledgement, PED (Pressure Equipment Directive) Compliance, a Declaration of Conformity, FAT (Factory Acceptance Test), a Drawing Schedule, ISO 9001:2008 Certification, Pressure vessel specification and door safety checks. The OQ Documentation includes an Automatic control test sheet (per cycle) and Chamber temperature distribution (per cycle)
Please note: This is our standard IQ/OQ Documentation package.

IQ/OQ

Assisted Air Cooling - REF: AAP007

A powerful external fan helps to reduce cool down time. The operation of the assisted air cooling system is controlled via the ‘Logi’ colour touch screen and is available on all models as an additional option, except for those fitted with a vacuum.

Assisted Cooling Option

Condensate Bottle (Classic Models Only) - REF: AAN308

The condensate bottle can be connected to any of the ‘Classic’ compact units to allow up to 4 litres of water collection during the freesteaming stage. This option is useful in situations where no easy drainage is available as a lower cost alternative to the ‘Autofill’ version(s).

Condensate Bottle

Five Position Shelf Rack - REF: AAN530

The 5 position shelf rack allows the user to vary what they are sterilizing in the autoclave by moving the positioning of internal shelves (supplied with this option). The AAN530 comes complete with 3 shelves supplied.

Astell Scientific Shelves

Spare Shelf - REF: AAN532

The spare shelf optinon for front loading Benchtop autoclaves.
Note: You must also purchase the 5 position shelf rack to be able to use this option.

Astell Scientific Shelves

Morrison Discard Container - REF: AAN346

Morrison discard containers allow the user to easily load batches of normally awkward shaped items.(Dimensions = 180(w) x 260(d) x 200(h) mm).

The UMB230BT can hold TWO AAN346 Discard Container.

Astell Discard Container

Contact Astell Inc





To enable us to quote you as quickly as possible, please let us know the following in your message:

  • Are you sterilizing any fluids?

  • Are you sterilizing any wrapped instruments or fabrics?

  • Are you sterilizing anything other than glassware, plastics or media?